Ultra-Technekow™ V4 Generator

Safety & Accuracy
&
Worldwide Performance

Features

The Ultra-Technekow™ V4 (technetium Tc 99m generator), referred to hereafter in this manual as the Ultra-Technekow V4 generator or UTK-V4 generator, may be received, used and administered only by authorized persons in designated clinical settings. Its receipt, storage, use, transfer and disposal are subject to the regulations and/or appropriate licenses of local competent official organizations. Users are to be nuclear medicine professionals who have had training in the safe use and handling of radioactive material and who also have received training specific to the nuclear pharmacy and these tools. Reading this manual will allow the operator to become familiar with components and procedures.
Intended Use
The Pharmacy Tools are used in the process of eluting sodium pertechnetate Tc 99m solution from the UTK-V4 generator. These tools are only to be used with Curium’s UTK-V4 generators.
Qualifications
Users are to be nuclear medicine professionals who have had training in the safe use and handling of radioactive material and also who have received training specific to the nuclear pharmacy and the tools. The tools are used typically in a controlled environment such as a nuclear pharmacy or nuclear medicine facility. The daily setup of the device is typically performed by trained nuclear medicine professionals.
Specifications
The Pharmacy Tool Pack consists of components that are intended to be used for both elution and dispensing of the radionuclide solution. The Auxiliary Shield Top, Elution Tool, Saline Shield, and Technestat™ Vial Holder are comprised of high density material that provides shielding protection to users during elutions. The Auxiliary Shield Cover is used to protect the Auxiliary Shield. The Tip Cap Replacement Tool is used to cover the elution and eluant needles with stored tip cap plugs. The Tip Cap Replacement Tool Ring and Technestat Vial Holder Ring are tool holders for the Tip Cap Replacement Tool and Technestat Vial Holder, respectively. The Saline Vial Alignment Insert adapts the eluant vial to the Saline Port of the Auxiliary Shield Top. Please note that the Elution Tool is shipped separately and is a component of the Pharmacy Tool Pack

Technical Specifications

ParameterDetail
IsotopeMolybdenum Mo-99 decaying to Technetium Tc-99m
Half-Life (Mo-99)66 hours
Half-Life (Tc-99m)6 hours
Gamma Emission Energy140.5 keV
Gamma Ray Constant0.795 R/hr-mCi at 1 cm
First Half-Value Layer0.023 cm lead
Shielding MaterialsLead, tungsten, or depleted uranium
Filter0.22 micrometer sterile filter
Maximum Mo-99 Activity19 Ci
Maximum Tc-99m ActivityDetermined by Mo-99 decay and elution interval